Performance Parameters:
Maximum Recommended Traffic (Flow Rate): 5 Liters per minute(L/min)
Oxygen Concentration: ≥90%(V/V) when the maximum recommended flow rate is 5L/min
Oxygen Concentration Indication (OSCI) Tolerances: Oxygen Concentration: ±3%; Flow rate: ±10% or 0.2L/min (take the maximum)
Maximum Outlet pressure during oxygen generation: 30-80kPa (error ± 10%)
Pressure Range during nebulization treatment: 55kPa~160kPa
Gas flowrate during atomization: 6L/min above
Concentration flow sensor operating temperature range: 5°C~50°C Atmospheric pressure range used by the flow sensor: 700hPa ~1060hPa
Noise: not more than 60dB(A)
Compressor relief valve operating pressure range: 250kPa ± 50kPa
Atomization Amount: 0.15mL/min
Power: AC 230V 50Hz
Rated Power: 380VA
Dimensions: 620mm X 375mm X 252mm
Electric Shock Protection: Class II Equipment, Type BF Applied Parts
Electromagnetic Compatibility: Group 1, Class B Equipment
Liquid Ingress Protection Classification: IPX0 Disinfection and Sterilization method: Follow manufacturer’s recommendations
Classification of operation modes: Continue to operate Safety Classification: Equipment that cannot be used in the presence of flammable anaesthetic gas mixtures with air or with oxygen or nitrous oxide.
Operation Temperature and Humidity: +5°C~40°C, ≤80% RH
Oxygen Generator Transportation and storage conditions: -25°C~+55°C(this product meets the requirements of GB/T 14710 standard for medium and low temperature storage(-40°C). In order to better ensure the stability of product performance, it is recommended that this product can be transported and stored at a temperature not lower than -20°C.) ≤93%RH(well ventilated, no corrosive gas.)
Operating atmospheric pressure 860kPa~1060kPa
Operating atmospheric pressure: 860kPa~1060kPa
Transport and storage atmospheric pressure: 500kPa~1060kPa
Period of use: 5 years