OX002 Oximeter Specifications:
1.Display Type: LED SpO2
Display range: 0%~100%. Pulse Rate Display range: 25bpm~250bpm.
2.Power: Two AAA 1.5V alkaline batteries.
3.Working Current: Less than 40mA at rated voltage 3V.
4.Measurement accuracy:
SpO2 Accuracy(Arms): ±2% in the range of 70%-100% of SpO2; No definition for SpO2 under 70%.
Pulse rate: 25bpm~250bpm, accuracy(Arms): ±1% or ±1bpm(larger)
Note: The accuracy(Arms) is calculated by the measurement values after a statistical distribution;compared to the reference device in a control study, approximately 2-thirds of the values were at(over or below) the accuracy(Arms) value.
5.Anti-interference ability of ambient light: Deviation in blood oxygen content is less than ± 1% when measured under indoor nature light / existing lighting and measured in the dark room.
6.The product will automatically shut down when there is no signal detected for about eight seconds.
7.Dimension: Length: 58mm Width: 33mm Height: 34mm Weight:36g approximately (without batteries).
8.Working Environments: Ambient temperature: 5℃~40℃ Relative humidity: ≤80% Atmospheric pressure: 860hPa~1060hPa.
9.Operation mode: Continuous operation.
10.Device response time.(See Figure 2)
11.Peak wavelengths and light output power: Emission wavelength range 600nm-1000nm, radiation intensity is less than 15mW/sr (20mA). Information of wavelength range may be of especial use to clinical doctors.
12.Description of the effect on displayed and transmitted SpO2 and Pulse rate data value:
1)data averaging and other signal processing.
2)the data update period:≤3 pulse rate cycles, less than 30s.
Note: Data processing and update will not affect pulse rate and SpO2.
13.The pulse waveform has been normalized, the measurement value is the best when the waveform is smooth and stable.
14. Internally Powered ME Equipment
15.TYPE BF APPLIED PARTS
16.Degrees of protection provided by enclosures (IP code):IP22.
E69 Thermometer Specifications:
Product Code: E69
Measuring Distance: ≤3cm
Measuring Site: Forehead/Object Surface
Measurement Range: Body Temperature : 32.0 ~ 43.0°C | Object temperature: 5.0 ~ 95.0°C
Accuracy: Body Temperature: ±0.3°C/±0.6°F | Object Temperature: ±1.0°C
Resolution: 0.1°C(0.1°F)
Reference Body Site: Axillia
Working Environment: (15~40)°C RH: (20~85)%RH
Storage Environment: (120~55)°C RH: ≤93%
Power Supply: DC 1.5V x 2(2 AAA 7#)
Memory: Previous temperature record
Display Screen: Large Screen